Covers Y25 to Y160, a total of 10 shell diameters. Select installation direction, shell and connector material, accuracy, scale and thread. The system automatically assembles the model code according to the naming rules, and simultaneously outputs the pressure unit conversion result.
Convert according to international unit standards and engineering definitions (1 psi = 6894.757 Pa, 1 kgf/cm² = 98066.5 Pa, 1 mmHg = 133.3224 Pa, 1 inH₂O = 249.0889 Pa).
Gauge pressureTaking atmospheric pressure as zero, all pressure gauge readings are gauge pressure;Absolute pressureTaking absolute vacuum as zero, absolute pressure ≈ gauge pressure + 101.325 kPa.
Dual scale: Inner circle bar, outer circle psi is a common practice. The "Scale" on this page also provides conversions for MPa / psi, which can be directly compared with the quotation and drawings.
kgf/cm²differs from bar by about 2%, these are two different units, not different ways of writing the same scale.
Values are all calculated from source values, no human rounding is done; engineering quotations and scale markings usually take 3-4 significant digits.
| Unit | Pa | kPa | MPa | bar | psi | kgf/cm² |
|---|
The bold "1" indicates that the unit in this row is itself. Water column units are calculated based on 4°C pure water density, and mercury columns are calculated based on 0°C mercury density, consistent with industry common practices for instrument rows.
| Scale (bar) | MPa | psi | kgf/cm² | kPa | Applicable specifications |
|---|
"Applicable specifications" automatically matches each specification scale sequence in the table; the scale system of Y25 and Y30 is obviously smaller than medium and large specifications, and small specifications do not produce high-pressure scales. This table is for accurate conversion values, and the dial scale is marked according to the integer rule in Section 04.
Letters are assembled in order, numbers are shell diameters, separated by "-" in between.
Dry type and oil-filled shockproof are mutually exclusive.When selecting dry type, oil or silicone oil cannot be selected; after selecting shockproof oil, dry type cannot be selected.
Iron shell and oil-filled shockproof are mutually exclusive.Iron shell (coated iron shell) can only be dry type; oil-filled shockproof requires a stainless steel shell.
Restrictions come from parameter table notes:
"When selecting dry type and iron shell, shockproof oil cannot be selected. After selecting shockproof oil, dry type or iron shell cannot be selected."
General engineering practice: static pressure scale is 1/3 to 3/4, dynamic or impact pressure is reduced by one level.After filling in "Actual working pressure", this page will provide real-time scale utilization ratio and recommendations. Less than 30% indicates that the scale is too large and the reading accuracy is wasted; more than 75% poses a risk of overpressure.Recommended range 30% – 75%
Dial value rounding
When "Custom scale / brand" or "Custom LOGO" is selected as "Yes", additional requirements must be filled in with at least 10 characters, otherwise the model order cannot be duplicated.