Covers Y25 to Y160, a total of 10 case diameters. Select installation direction, case and connection material, accuracy, scale and thread pitch. The system automatically assembles the model code according to the naming rules, and simultaneously outputs the pressure unit conversion result.
Convert according to international unit standards and common engineering definitions (1 psi = 6894.757 Pa, 1 kgf/cm² = 98066.5 Pa, 1 mmHg = 133.3224 Pa, 1 inH₂O = 249.0889 Pa).
Gauge pressureTaking atmospheric pressure as zero, all pressure gauge readings are gauge pressure;Absolute pressureTaking absolute vacuum as zero, absolute pressure ≈ gauge pressure + 101.325 kPa.
Dual scale: Inner scale bar, outer scale psi is a common practice. This page's 'Scale' also provides MPa / psi conversions, which can be directly referenced for quotation and drawings.
kgf/cm²differs from bar by about 2%, these are two different units, not different ways of writing the same scale.
Values are all converted from source values, and no human rounding is done; engineering quotes and scale markings usually take 3-4 significant digits.
| Unit | Pa | kPa | MPa | bar | psi | kgf/cm² |
|---|
Bold '1' indicates the unit of this row. Column units are converted based on 4 °C pure water density, and mercury columns are converted based on 0 °C mercury density, consistent with industry common practices.
| Scale (bar) | MPa | psi | kgf/cm² | kPa | Applicable specifications |
|---|
'Applicable specifications' automatically matches each specification scale sequence in the table; Y25 and Y30 have significantly smaller scale systems than medium and large specifications, and small specifications do not produce high-pressure scales. This table is for accurate conversion values, and the dial scale marking follows the rounding rules in Section 04.
Letters are assembled in order, numbers are case diameters, separated by '-' in between.
Dry type and oil-filled are mutually exclusive.When selecting dry type, oil-filled or silicone oil cannot be selected; after selecting shockproof oil, dry type cannot be selected.
Iron case and oil-filled are mutually exclusive.Iron case (coated iron case) can only be dry type; oil-filled shockproof requires a stainless steel case.
Exclusions come from parameter table notes:
'Select dry type, iron case cannot select shockproof oil; after selecting shockproof oil, dry type or iron case cannot be selected.'
General engineering method: Static pressure is the scale of 1/3 ~ 3/4, pulse or impact pressure is reduced by one level.
After filling in 'Actual working pressure', this page will provide real-time scale utilization ratio and recommendations; less than 30% indicates the scale is too large, wasting reading accuracy; greater than 75% poses an overpressure risk.
When selecting bar, keep the value in the table; when selecting MPa, take 1/10 of the original value; when selecting psi, take 14.5 times the original value and round it, keeping at most three significant digits, and append zeros at the end.
Example: 15 bar → 1.5 MPa → 218 psi (15 × 14.5 = 217.5, rounded to three significant digits is 218).